Priority
Icon
Level
General Fault Codes
Notes:
• If more than one fault is present, the device status indicator and text for each fault
alternate at 2 second intervals.
• Some faults cause the device to go to Fail-safe mode (Fault 52). These are indicated with
an asterisk (*).
Code /
Meaning
Icon
S: 0
*
The device was unable to get a mea-
surement within the Fail-safe LOE
Timer period. Possible causes: faulty
installation, antenna material
buildup, foaming/other adverse pro-
cess conditions, invalid calibration
range.
S: 2
*
Unable to collect profile because of
a power condition that is outside of
the operating range of the device.
S: 3
Device is nearing its lifetime limit
according to the value set in Mainte-
nance Required Limit.
S: 4
Device is nearing its lifetime limit
according to the value set in Mainte-
nance Demanded Limit.
A5E32337685
SITRANS LR260 (PROFIBUS PA) – OPERATING INSTRUCTIONS
Meaning (cont' d )
• Data exchanged
• No data exchange
• Write access enabled
• Write access disabled
General Fault Codes
Corrective Action
Ensure installation details are cor-
rect. Ensure no antenna material
buildup. Clean if necessary. Adjust
process conditions to minimize foam
or other adverse conditions. Correct
range calibration. If fault persists,
contact your local Siemens repre-
sentative.
Repair required: contact your local
Siemens representative.
Replacement is recommended.
Replacement is recommended.