List Of Error Numbers - Siemens SITRANS FM MAG 5000 Operating Instructions Manual

Electromagnetic flowmeters ip67
Hide thumbs Also See for SITRANS FM MAG 5000:
Table of Contents

Advertisement

9.2

List of error numbers

Error text
Error
No.
Remedy text
I1 - Power on
1
OK
I2 - Add-on module
2
Applied
I3 - Add-on module
3
Install
I4 - Param. corrected
4
OK
W20 - Totalizer 1
20
Reset manually
W20 - Totalizer 2
20
Reset manually
W21 - Pulse overflow
21
Adj. pulse settings
W22 - Batch timeout
22
Check installation
W23 - Batch overrun
23
Check installation
W24 - Batch neg. flow
24
Check flow direction
W30 - Overflow
30
Adj. Q
max
W31 - Empty pipe
31
P40 - SENSORPROM®
40
Insert/change
P41 - Parameter range
41
Switch off and on
SITRANS FM MAG 5000/6000 IP67
Operating Instructions, 12/2019, A5E02338368-AD
Comment
Device powered on
A new module has been applied to the system
An add-on module is defect or has been removed.
This can be an internal add-on module
A less vital parameter in the transmitter has been re‐
placed by its default value
During initialisation the check of the saved totalizer
value has failed. It is not possible to rely on the saved
totalizer value anymore. The totalizer value must be
reset manually in order to rely on future readings
During initialisation the check of the saved totalizer
value has failed. It is not possible to rely on the saved
totalizer value anymore. The totalizer value must be
reset manually in order to rely on future readings
Actual flow is too big compared with pulse width and
volume/pulse
Duration of batching has exceeded a predefined max‐
imum time
Batch volume has exceeded a predefined maximum
overrun volume
Negative flow direction during batch
Flow is above Q
settings
max
Pipe is empty
SENSORPROM
unit not installed
®
A parameter is out of range. The parameter could not
be replaced by its default value. The error will disap‐
pear at the next power-on
Diagnostics and Troubleshooting
9.2 List of error numbers
Output
status
Active
Active
Active
Active
Active
Active
Reduced
pulse width
Batch output
on zero
Batch output
on zero
Active
Max. 120 %
Zero
Active
Active
Input
status
Active
Active
Active
Active
Active
Active
Active
Active
Active
Active
Active
Active
Active
Active
63