Recording Ise Slopes - Siemens ADVIA 1800 Operator's Manual

Chemistry system
Table of Contents

Advertisement

NOTE:
part of the shutdown wash and twice on a per-shift basis). Pour fresh ISE Detergent
into a cup, not a tube, before each wash from the CTT.
1. At the Menu Panel, select
2. At the ISE Operation window, in the Period.wash area, select
Set
3. In the Wash Electrode area do the following:
a. Place a cup of the ISE Detergent in the identified position on the screen.
b. In the Container field, select the type of container for the wash solution.
4. In the Wash Electrode area, select
5. At the ISE Operation window, in the Period.wash area, select
Set
6. Close the window, then select

Recording ISE slopes

Once a day, record the slopes from a successful ISE calibration on the Maintenance Log.
The slopes are provided on the ISE Monitor, RBL/Calibration History, and RealTime
Monitor windows following a successful calibration.
130
Do not perform the ISE electrode wash more than 3 times per day (once as
.
The recommended type of container is 6 : 2mlCUP/Adp.
.
., then select
Maint
ISE Operation
, then select
Execute
when prompted.
Yes
.
, then select
OFF
when prompted.
Yes
, then select
ON
ADVIA 1800 Operator's Guide