3. Loosen but do not remove the probe wire screw (
wire from the post.
4. Hold the probe arm securely and open the two spring clips (
side closest to the black wire (
5. Loosen but do not remove the wire lock screw (
6. Remove the black wire (
7. Gently snap the gold joint holder (
8. Gently lift the probe (
the probe arm.
9. Discard the old probe.
Install the new DPP probe
1. Carefully insert the new probe into the probe guide (
the joint connector (
2. Press the joint end of the probe into the joint holder (
3. Carefully close each spring clip (
4. Reconnect the black wire (
5. Slip the orange wire onto the probe wire screw (
6. Carefully flex the tubing (
7. Replace the probe arm cover and tighten the two probe cover screws.
Maintenance
5
CAUTION
There is some spring resistance when attempting to open the clips. Do not
allow the probe arm to swing side to side when opening the clips.
) from the post.
5
8
) up through the probe guide (
) into the back end of the probe and carefully tighten until snug.
2
CAUTION
Do not cross thread or force the joint connector in too far, to avoid damaging
the threads or introducing leaks or air bubbles.
CAUTION
Do not allow the clips to snap on the probe shaft. This may cause damage to
the probe.
) under the wire lock screw (
5
CAUTION
Do not flex the wire more than necessary. Over-flexing may damage the wire.
CAUTION
If the screw does not fully tighten, or the standoff spins, tighten the screw on
the probe arm base until the standoff no longer spins; otherwise the liquid-
level-sensing capability may be adversely affected.
) and slip the flared end into the joint connector (
1
9
), then remove the orange probe
), then gently raising each to an open, locked position.
).
9
3
) from its post.
7
), then carefully remove it from
7
), then slide the knurled nut of
3
).
) over the probe.
6
), then tighten the screw.
9
) by grasping each at the
6
) and tighten the screw.
4
).
2
165