2. Maintenance work
A screen for use in the maintenance work by a service personnel is described.
2.1 Setting of maintenance mode by service personnel
For the maintenance work by a service personnel, restrictions on use of the following
functions are removed.
●System environment setup
●Adjustment/operational checkout
●Solenoid valve/pump test
●Download
Maintenance menu
Environment setup
menu
Excitation update
(Nozzle test)
Software management
1 Select the maintenance in a menu and then select Password protection.
Password protection
After you have entered a correct password, you are permitted to
change the password protection settings below.
Password [ ]
Edit message/Select message [0]
Save message
Print specifications
Print format
Maintenance
2 Input a password "HALB" and press Limitation change.
[Caution] 1. The password "HALB" is a special password to enable the
predetermined screen related to the control of IJP. Do not tell this
password to the customer.
2. Even when the customer sets the password in the functional
restriction, the "HALB" has priority.
Status:XXXXX
Operation
Test print
management
Log management
Circulation control
Password
protection
Status:XXXXX
(0:access
1:protect)
[0] (0:access 1:protect)
[0]
(0:access
1:protect)
[0]
(0:access
1:protect)
[0]
(0:access
1:protect)
2-1
Shut
Manual
down
Back
Limitation
change
Back