2.3.5 Level sensor confirmation
●The operation confirmation of level sensor in the main ink vessel is performed.
①Level over
●
①Insufficient ink
●
①Insufficient makeup ink
●
With respect to the
password protection,
refer to " 2.1 setting of
service personnel
maintenance mode"
1 Press Adjustment/operational checkout at the second page of the service
2 Press Level sensor confirmation.
3 When pressing Back, the screen returns to the adjustment/operational
②Initial level
The operation confirmation of level sensor in the ink vessel is performed.
The operation confirmation of level sensor in the makeup ink vessel is performed.
Perform this test in a "Stop" or "Standby" state.
Maintenance menu
System
environment setup
operational checkout
Download
Unit information
maintenance screen.
When the Next menu are not displayed in the maintenance work menu, set to
the service personnel maintenance mode by the password protection.
Adjust./oper. checkout
Excitation
adjustment
Auto-phase gain
adjustment
The level sensor operation state is displayed.
Level sensor confirmation
<Main ink reservoir>
Too high [ detected undetected]
Initial [ detected undetected]
<Ink reservoir>
Too low [ detected undetected]
<Makeup ink reservoir>
Too low [ detected undetected]
Detected
: The sensor detected the level.
(The ink and the makeup ink are present.)
Not detected : The sensor did not detect the level.
(The ink and the makeup ink are not present.)
checkout screen.
Status:XXXXX
Adjustment/
Solenoid valve/
pump test
Language change
System reset
Status:XXXXX
Charge voltage
Phase margin test
confirmation
Level sensor
Heating unit
confirmation
confirmation
Status:XXXXX
2-12
Shut
Manual
down
Next
menu
Back
Shut
Manual
down
Back
Back