hit counter script

Connecting The Stand Cable-Harness; High-Voltage Connector - Siemens MAMMOMAT 1000 Nova Installation And Start-Up Instructions Manual

Hide thumbs Also See for MAMMOMAT 1000 Nova:
Table of Contents

Advertisement

5 - 4

Connecting the stand cable-harness

High-voltage connector

Fig. 4 High-voltage connector
NOTICE
M1000/3000 Nova
Installation and Start-Up
3
1
2
   1
   6
   1
   0
   0
   M
   A
   M
The high-voltage connector (1) is factory-adjusted for correct fit.
The high-voltage connector shall be inserted with silicone oil, sili-
cone-rubber ring (2) and silicone-rubber disc (3). Lubricate the sil-
icone-rubber disc and the pins with silicon oil before mounting
the disc. Make sure that there aren't any air bubbles between the
disc and the connector.
Register 3
SPB7-230.033.11
Rev. 06
Cable connections
Page 4 of 14